Abstract
In the present study, beam quality correction, [Formula: see text], and phantom scatter correction, kphan(r), for low-energy brachytherapy sources, 131Cs, 125I, and 103Pd, are calculated using the Monte Carlo-based EGSnrc code system as a function of the distance along the transverse axis of the source. The solid-state detectors investigated are diamond, LiF, Li2B4O7, Al2O3, and radiochromic films, such as HS, EBT, EBT2, EBT3, RTQA, XRT, and XRQA. The solid phantoms investigated are polystyrene, PMMA, virtual water, solid water, plastic water (LR), A150, RW1, RW3, and WE210. For a given detector and brachytherapy source, [Formula: see text] is independent of distance in the water phantom. Meanwhile, for a given detector, kphan(r) depends on the distance from the source for the investigated solid phantoms. Moreover, the kphan(r) values do not change with the detector type for sources 131Cs, 125I, and 103Pd at all distances. The LR and A150 phantoms are water equivalent for the investigated distances of 1-5cm. The phantoms including solid water, virtual water, and WE210 are not water-equivalent for distances beyond 1cm. Furthermore, PMMA, polystyrene, RW1, and RW3 are not water equivalent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.