Abstract

Multiple drug target analysis (MDTA) used in doping control is more efficient than single drug target analysis (SDTA). The number of drugs with the potential for abuse is so extensive that full coverage is not possible with SDTA. To address this problem, a liquid chromatography tandem mass spectrometric method was developed for simultaneous analysis of 302 drugs using a scheduled multiple reaction monitoring (s-MRM) algorithm. With a known retention time of an analyte, the s-MRM algorithm monitors each MRM transition only around its expected retention time. Analytes were recovered from plasma by liquid-liquid extraction. Information-dependent acquisition (IDA) functionality was used to combine s-MRM with enhanced product ion (EPI) scans within the same chromatographic analysis. An EPI spectrum library was also generated for rapid identification of analytes. Analysis time for the 302 drugs was 7 min. Scheduled MRM improved the quality of the chromatograms, signal response, reproducibility, and enhanced signal-to-noise ratio (S/N), resulting in more data points. Reduction in total cycle time from 2.4 s in conventional MRM (c-MRM) to 1 s in s-MRM allowed completion of the EPI scan at the same time. The speed for screening and identification of multiple drugs in equine plasma for doping control analysis was greatly improved by this method.