Abstract

Based on an efficient sample clean-up and field-amplified sample injection online preconcentration technique in capillary electrophoresis with contactless conductivity detection, a new analytical method for the sensitive determination of melamine in milk samples was established. In order to remove the complex matrix interference, which resulted in a serious problem during field-amplified sample injection, liquid-liquid extraction was utilized. As a result, liquid-liquid extraction provides excellent sample clean-up efficiency when ethyl acetate was used as organic extraction by adjusting the pH of the sample solution to 9.5. Both inorganic salts and biological macromolecules are effectively removed by liquid-liquid extraction. The sample clean-up procedure, capillary electrophoresis separation parameters and field-amplified sample injection conditions are discussed in detail. The capillary electrophoresis separation was achieved within 5min under the following conditions: an uncoated fused-silica capillary, 12mM HAc + 10mM NaAc (pH = 4.6) as running buffer, separation voltage of +13kV, electrokinetic injection of +12kV × 10s. Preliminary validation of the method performance with spiked melamine provided recoveries >90%, with limits of detection and quantification of 0.015 and 0.050mg/kg, respectively. The relative standard deviations of intra- and inter-day were below 6%. This newly developed method is sensitive and cost effective, therefore, suitable for screening of melamine contamination in milk products.