Abstract

The power management section associated with the biomedical circuit is very crucial and should be optimally designed for optimal utilization of power. This work discusses the different power shaping or conversion circuits that had been taken for their performance analysis. The two-performance metrics power conversion efficiency and susceptibility against the wireless power transfer have been taken to investigate the operational performance of the biomedical circuits against the input signal strength and operating frequencies. Simulated results confirm the CNFET-based circuit performance is very good at a small value of input voltage 0.6V and a broad range of operating frequency (953 MHz). Therefore, a CNFET-based circuit can be used suitably in implantable devices with optimum power utilization and a remote powering mechanism over the RF link.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call