Abstract
A simple process for realizing stable and reliable electrospray ionization (ESI) tips in polymer microfluidic systems is described. The process is based on the addition of a thin hydrophobic membrane at the microchannel exit to constrain lateral dispersion of the Taylor cone formed during ESI. Using this approach, ESI chips are shown to exhibit well-defined Taylor cones at flow rates as low as 80 nL min(-1) through optical imaging. Furthermore, stable electrospray current has been measured for flow rates as low as 10 nL min(-1) over several hours of continuous operation. Characterization of the electrospray process by optical and electrical monitoring of fabricated ESI chips is reported, together with mass spectrometry validation using myoglobin as a model protein. The novel process offers the potential for low-cost, direct interfacing of disposable polymer microfluidic separation platforms to mass spectrometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.