Abstract

The circle system has been in use for more than a 100 years, whereas the first clinical application of an anaesthetic reflector was reported just 15years ago. Its functional basis relies on molecular sieves such as zeolite crystals or activated carbon. In a circle system, the breathing gas is rebreathed after carbon dioxide absorption; a reflector on the other hand specifically retains the anaesthetic during expiration and resupplies it during the next inspiration. Reflection systems can be used in conjunction with intensive care ventilators and do not need the permanent presence of trained qualified staff. Because of easy handling and better ventilatory capabilities of intensive care ventilators, reflection systems facilitate the routine use of volatile anaesthetics in intensive care units. Until now, there are three reflection systems commercially available: the established AnaConDa™ (Sedana Medical, Uppsala, Sweden), the new smaller AnaConDa-S™, and the Mirus™ (Pall Medical, Dreieich, Germany). The AnaConDa consists only of a reflector which is connected to a syringe pump for infusion of liquid sevoflurane or isoflurane. The Mirus represents a technical advancement; its control unit includes a gas and ventilation monitor as well as a gas dispensing unit. The functionality, specific features, advantages and disadvantages of both systems are discussed in the text.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.