Errors are likely to occur due to defects, chip degradation, and the lack of precision inherent in biochemical experiments. Therefore, an efficient error-recovery strategy is essential to ensure the correctness of assays executed on micro-electrode-dot-array (MEDA) biochips. By exploiting MEDA-specific advances in droplet sensing, this chapter presents a novel error-recovery technique to dynamically reconfigure the biochip using real-time data provided by on-chip sensors. Local recovery strategies based on probabilistic-timed-automata are presented for various types of errors. An online synthesis technique and a control flow are also proposed to connect local-recovery procedures with global error recovery for the complete bioassay. Moreover, an integer linear programming-based method is also proposed to select the optimal local-recovery time for each operation. Laboratory experiments using a fabricated MEDA chip are used to characterize the outcomes of key droplet operations. The PRISM model checker and three benchmarks are used for an extensive set of simulations. Our results highlight the effectiveness of the proposed error-recovery strategy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.