Abstract

Monte Carlo calculations of a bismuth-germanate scintillator's efficiency agree closely with experimental measurements. For this comparison, we studied the absolute gamma-ray photopeak efficiency of a scintillator (7.62 cm long by 7.62 cm in diameter) at several gamma-ray energies from 166 to 2615 keV at distances from 30.5 to 152.4 cm. Computer calculations were done in a three-dimensional cylindrical geometry with the Monte Carlo coupled photon-electron code CYLTRAN. For the experiment we measured 11 sources with simple spectra and precisely known strengths. The average deviation between the calculations and the measurements is 3%. Our calculated results also closely agree with recently published calculated results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.