Abstract

Laboratory testing for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has played an important role in the effort to prevent and contain local outbreaks. The aim of this study was to assess the diagnostic accuracy of a new fully automated SARS-CoV-2 laboratory-based antigen assay (CoV2Ag) and to explore the efficiency of a diagnostic algorithm combining antigen and conventional high-throughput molecular assays to address potential future challenges of the SARS-CoV-2 pandemic. One thousand two hundred and twenty four consecutive nasopharyngeal swabs were tested using RT-PCR and CoV2Ag assay. The overall sensitivity and specificity of CoV2Ag were 79.1 and 97.8%, respectively. When the analysis wasrestricted to cases with Ct values≤30, the sensitivity of the assay improved to 98.1%. Acceptable sensitivity was found when the analysis was limited to patients presenting within one or two to four days of symptom onset (80.5 and 84.8%, respectively). A retrospective analysis of the use of a two-step diagnostic approach combining the CoV2Ag assay and RT-PCR during an acute pandemic phase of 97days showed a potential reduction in the number of RT-PCR tests by 36.1%, corresponding to savings in reagent costs and technician workload of approximately €8,000 and 10.5h per day, respectively. Our data show that the proposed algorithm represents a valid alternative diagnostic approach to increase testing efficiency during future pandemic phases with high positivity rates (>20%) and elevated numbers of RT-PCR test requests.