Aims To determine the efficacy of super-oxidised water (Optident/Sterilox) in the decontamination of dental unit water lines. Methods Dental units (10) were first purged with concentrated super-oxidised water. After purging, a 5% (v/v) super-oxidised water was used as a maintenance dose. Samples for microbiology were taken after 0,1,2,3,4,5,6,7 d, and each week for a further 13 weeks. Results After purging, 5% (v/v) super-oxidised water was successful in reducing the microbial counts to zero, although in three of the units some bacteria were intermittently isolated in the first week of treatment. Conclusions Super-oxidised water was successful in the removal of bacteria from dental unit water supplies. Complete removal required the treatment with a purge phase of concentrated disinfectant and a maintenance phase of at least two weeks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.