Abstract

This study aimed to use optical frequency domain imaging (OFDI) to evaluate the efficacy of post-dilatation (PD) after stent placement for unstable plaques during carotid artery stenting (CAS) using a double-layer stent. Twelve unstable carotid plaque lesions diagnosed by MRI were evaluated using OFDI during CAS. The pre-procedural minimum lumen diameter was 1.6 ± 0.7 mm. Each lesion was pre-dilated with balloon catheters (diameter, 5.3 ± 0.5 mm), and a double-layer stent was deployed. PD was performed with balloon catheters of the same size as those used for pre-dilatation. Cross-sectional OFDI images within the stented segment were evaluated at 1-mm intervals for a 20-mm segment, including the most stenotic lesion. Slice rates for the presence of in-stent plaque protrusion (PP) and plaque between the double-layer lumen were calculated. No procedural complications occurred with the use of an embolic protection device. Compared to after stent placement, slice rates for any PP (44 ± 19% to 62 ± 22%, P <0.05) and plaque between the double-layer lumen (79 ± 16% to 91 ± 34%, P <0.05) were significantly increased after PD; slice rates for >500 μm PP (7.5 ± 14% to 0%, P <0.05) were significantly decreased. Visible debris were captured in 50% of lesions. PD after double-layer carotid stent placement decreases in-stent large PP. Double-layer construction contributed to the prevention of large PP, as the PP may have been crushed into debris by PD.