Abstract

To assess effectiveness of an experimental sterilization method based on the exposure of an O3/O2 gas mixture directly inside the packaging for clear aligners. Fifty samples consisting of pieces of polyethylene terephthalate glycol (PET-G) aligners were contaminated by manual handling and subsequently divided into different groups (n = 30 for exposure to O3/O2 gas at different times, n = 10 for positive control with 2% chlorhexidine digluconate, n = 10 for negative control). The measurement of optical densities (OD) of the initial and final microbial cultures was recorded for all groups. Kruskal-Wallis test was used for differences between groups while Wilcoxon test was used to compare initial and final OD values within groups. Statistical significance was set at P < .05. Comparison within the groups showed statistically significant differences for exposure to the gaseous mixture (72 hours), for positive and negative controls. Other significant differences were found in the multiple comparisons between the application of gaseous ozone (48 hours and 72 hours) and the negative control. The direct exposure of gaseous ozone on the aligners inside their packaging showed microbicidal capacity at 72 hours, which was equivalent to the positive control with immersion in chlorhexidine digluconate. This innovative sterilization procedure could be considered in the final manufacturing processes of clear aligners to eliminate the potentially pathogenic microorganisms that are deposited on surfaces of these orthodontic devices.