Abstract

A newly designed hygroscopic condenser humidifier was evaluated to assess efficacy of providing moisture to the airways in patients whose upper airways had been bypassed. With its use, the relative humidity of inspired gas fell with both increasing tidal volumes and especially increasing F10(2) due to the intrinsic dryness of medical oxygen. Under conditions other than high volumes and inspired oxygen, humidification appeared adequate. The device obviated many complications encountered with conventional heated nebulizers. It does add some degree of resistance in the ventilatory circuit which might be stressful during spontaneous breathing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.