Abstract

To determine whether the use of a metal artifact reduction (MAR) algorithm improves the detection of peri-implant fenestrations and dehiscences on cone beam computed tomography scans. One hundred titanium fixtures were implanted into bovine ribs after the creation of defects simulating fenestrations and dehiscences. Images were acquired using four different protocols, namely, A2 (MAR on, voxel 0.2mm), A3 (MAR on, voxel 0.3mm), B2 (MAR off, voxel 0.2 mm), and B3 (MAR off, voxel 0.3mm). For all protocols, receiver operating characteristic (ROC) curves were determined. Values for the areas under the ROC curves (Az) were subjected to analysis of variance. Az values were not statistically different among protocols regardless of the defect type (P > .05). The MAR algorithm tested by us did not improve the diagnosis of peri-implant fenestrations and dehiscences with use of either the 0.2mm or the 0.3mm voxel sizes.