Abstract

ObjectiveTo observe the efficacy and safety of self-made covered coronary stent in the treatment of coronary artery perforation.MethodsCovered coronary stent was prepared by wrapping 3 M film on the surface of coronary stents. The beagle dogs were divided into control group and experimental group. A drug-eluting stent (DES) was implanted in the control group. The covered stent was applied to block the coronary branches of beagle dogs. The CaIMR value after stent placement was calculated by FlashAngio software. The effect of blocking the coronary branches on blood flow was observed by coronary angiography (CAG). The condition of the implanted stent was observed by optical coherence tomography (OCT), and the histopathologic examination of the coronary vessel implanted stent was performed by HE staining.ResultsThe best number of layers was 2. Compared with the control group, the CaIMR of the experimental group increased (p < 0.05). A lot of in-stent thrombosis were found in the experimental group and obvious blood flow obstruction during follow-up. HE staining showed that stents implanted in the two groups adhered well to the wall of the blood vessel, but in-stent thrombosis and intimal hyperplasia were founded in the experimental group, while the in-stent restenosis was not founded.ConclusionThe self-made coronary covered stent can effectively block the leakage caused by coronary perforation, but the stent endothelialization is poor, which easily causes stent thrombosis and restenosis, so it is not recommended as a routine remedy.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call