Abstract

Compared with currently available duodenoscopy-assisted systems, direct peroral cholangioscopy (DPOC) using an ultra-slim endoscope is limited by technical difficulties. The multibending (MB) ultra-slim endoscope was introduced as a dedicated cholangioscope for DPOC to challenge the technical problem. We retrospectively analyzed the clinical utility of DPOC using an MB endoscope with free-hand insertion into the bile duct in patients with biliary diseases. A total of 145 patients who underwent DPOC using an MB endoscope were analyzed. The primary outcome was the technical success rate of DPOC using the free-hand insertion of the MB endoscope. The secondary outcomes were the technical success rates of DPOC-guided diagnostic and therapeutic interventions, the diagnostic accuracy of DPOC-guided target biopsy, and adverse events related to DPOC. Free-hand biliary insertion of a MB endoscope for DPOC was technically successful in 133 patients (91.7%). DPOC-guided target biopsy was successful in 36 of 38 patients (94.7%) and had a diagnostic accuracy of 91.7% (95% confidence interval, 82.6-100). Sixty-nine therapeutic interventions were performed; technical success was achieved in 65 (94.2%). No severe adverse events were observed. The MB ultra-slim endoscope was technically effective to perform a DPOC including various diagnosis and therapeutic interventions without device assistance. MB endoscope is considered to contribute to expanding a role of DPOC in diagnosis and treatment of diverse biliary tract diseases.