Abstract

A compact injector for a heavy-ion medical-accelerator complex was developed. It consists of an electron-cyclotron-resonance ion-source (ECRIS) and two linacs, which are a radio-frequency-quadrupole (RFQ) linac and an Interdigital H-mode drift-tube-linac (IH-DTL). Beam acceleration tests of the compact injector were performed, and the designed beam quality was verified by the measured results, as reported earlier. Because the method of alternating-phase-focusing (APF) was used for beam focusing of the IH-DTL, the motion of beam ions would be sensitive to gap-voltage errors, caused during tuning of the gap-voltage distribution and by automatic-frequency tuning in actual operation. To study the effects of voltage errors to beam quality, further measurements were performed during acceleration tests. In this report, the effects of voltage errors for the APF IH-DTL are discussed.