Abstract

Several authors have reported promising results with vacuum-assisted closure therapy in poststernotomy mediastinitis. The aim of this study was to investigate the hemodynamic outcome following the application of six negative pressures on an open sternotomy wound. Six 70-kg pigs underwent median sternotomy followed by vacuum-assisted closure therapy. Six negative pressures (-50, -75, -100, -125, -150, and -175mmHg) were applied to each pig for 30min each while hemodynamic parameters were measured. An increase in cardiac output was observed at -75mmHg when compared to the other five pressures: -50mmHg (P<0.05; CI 0.12-1.13l/min), -100mmHg (P<0.001; CI 0.34-1.32l/min), -125mmHg (P<0.001; CI 0.51-1.52l/min), -150mmHg (P<0.001; CI 0.50-1.47l/min), and -175mmHg (P<0.05; CI 0.13-1.17l/min). A decrease in systemic vascular resistance was observed at -75mmHg when compared to -125mmHg (P<0.01; CI 108-552dyn.s/cm(5)) and -150mmHg (P<0.01; CI 90-543dyn.s/cm(5)), but not compared to the other pressures. No change (P=ns) was observed in heart frequency, mean arterial pressure or central venous pressure. Our data demonstrates that vacuum-assisted closure therapy of -50 to -175mmHg does not impair the central hemodynamics in a porcine sternotomy model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.