Diffuse optical imaging can be used to probe highly scattering media like biological tissue down to a depth of few centimeters, with spatial resolution limited by light scattering. Its combination with ultrasound imaging can potentially lead to medical imaging systems with, for instance, high specificity in the examination of tumors. However, the presence of the ultrasound coupling gel between probe and tissue can have detrimental effects on the accuracy of optical imaging techniques. Here we present an experimental study on the effect of ultrasound coupling fluids on diffuse optical spectroscopy (DOS) and diffuse correlation spectroscopy (DCS). We demonstrate on tissue-mimicking phantoms that the use of standard water-clear gels, providing a direct path for the light from the source to the detection point, can distort optical measurements generating strong underestimation of both the absorption and the reduced scattering coefficients in DOS measurements, as well as underestimation of the Brownian diffusion coefficient in DCS measurements. On the contrary, various turbid fluids demonstrate excellent performance in preventing this issue.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.