Abstract

ObjectiveTo investigate the influence of implant design factors in terms of bone integrity and implant stability. Materials and methodsA 3D parametric CAD model was developed. Then, once domain settings and boundary conditions were defined, a 3D FEM model was created. To simulate the physical interaction at the bone–implant interface, identity pairs were introduced. After generating different design scenarios with a DOE approach, the most significant design factors were obtained. ResultsThis study showed that the geometry of the screw thread highly influenced the implant stability. In particular the degree of bone damage became minimal when adopting 0.40mm for the thread width and 0.05mm for the thickness. SignificanceThread width and thickness play a crucial role to reduce induced stresses and damage in bone. Considering these preliminary results, future improvements should focus on investigating also two-factor and higher interactions to better understand the implant loading mechanism.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.