Abstract
We performed phantom experiments to investigate radiation dose in the computed tomography component of oncology positron emission tomography/computed tomography in relation to the scan range. Computed tomography images of an anthropomorphic whole-body phantom were obtained from the head top to the feet, from the head top to the proximal thigh or from the skull base to the proximal thigh. Automatic exposure control using the posteroanterior and lateral scout images offered reasonable tube current modulation corresponding to the body thickness. However, when the posteroanterior scout alone was used, unexpectedly high current was applied in the head and upper chest. When effective dose was calculated on a region-by-region basis, it did not differ greatly irrespective of the scan range. In contrary, when effective dose was estimated simply by multiplying the scanner-derived dose-length product by a single conversion factor, estimates increased definitely with the scan range, indicating severe overestimation in whole-body imaging.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.