Abstract

In recent years the use of devices called Heat and Moisture Exchangers (HME) has become widespread as gas conditioners for ICU patients requiring mechanical ventilation. As an important variation of the resistive properties of the HME, related to flow and duration of use, has recently been pointed out during "in vitro" studies, the use of these devices in COPD patients could increase the levels of auto PEEP and dynamic hyperinflation. In this study we have compared the levels of auto PEEP and difference in functional residual capacity (delta FRC) in a group of COPD patients, requiring controlled mechanical ventilation (CMV), at basal conditions and after the insertion into the circuit of three HMEs (Dar Hygrobac, Pall Ultipor, Engstrom Edith) at random: the results obtained excluded a significant increase of auto PEEP and delta (FRC) both with "new" HMEs and after 12 h of continuous use.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.