Abstract

Abstract The environment parameters, i.e. temperature, pressure, humidity and CO2 concentration affecting the refractive index of air play an important role on the long gauge block measurement using interferometer. In this paper, the environment parameters in a measurement chamber had been continuously investigated more than 24 h. The refractive index was then calculated by using an updated Edlen equation. The stability of the refractive index of air in the experiment is 0.1 × 10−6 with the measurement uncertainty of 4.32 × 10−8 at the confidential level k = 2. While the sensitivity of the pressure of air was small, it was an evidence that the pressure of air mainly contributed to the stability of the refractive index of air due to its large variation. Furthermore, the calculated refractive index of air was employed to correct the length of long gauge blocks. The correction for 1 m long gauge block in interferometer system is 0.270 mm, 0.269 mm and 0.267 mm which are respectively based on the three laser wavelengths at 532 nm, 633 nm and 780 nm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.