Abstract
A superconducting gravimeter based on the superconducting quantum interference device system is under development. As the main source of low-frequency noise, temperature fluctuations affect the resolution of superconducting gravimeters. In this study, a set of experimental devices was built to investigate the primary coupling processes of temperature fluctuations in superconducting gravimeters. Under the temperature modulation method, the effects of temperature fluctuations can be expressed as dΦ/dT = 342(2)Φ0/K, which, according to theoretical analysis, corresponds to a displacement change of (1.38 ± 0.04) × 10-7 m/K. Based on these results, the ambient temperature is controlled to within ±100 µK, and the equivalent effect of temperature fluctuations on our superconducting gravimeter is 0.5 μGal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.