Abstract

We report a single-beam synthetic gradiometer operated in the spin-exchange-relaxation free (SERF) regime, using the structure of two separate atomic vapor cells spaced 2 cm apart. To improve the capability of the gradiometer in suppressing the common-mode magnetic field noise, we are aiming at investigating the effects of the system parameters on the gradiometer common-mode rejection ratio (CMRR). The mathematical expression for the relationship between the gradiometer CMRR and the two variables including the linewidth ratio and the pumping factor ratio is constructed for the first time, to our knowledge. This means that the CMRR can be optimized by controlling the linewidth and the pumping factor, which is easy to implement in the operation process. As a result, a CMRR of 246 is achieved and a gradiometer sensitivity of 4.5 fT/cm/Hz1/2 is also measured. This method provides a theoretical and experimental basis for the automated operation of gradiometers, and the gradiometer system performance can be tuned to a desired state by simply controlling the linewidth and the incident light intensity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.