Abstract

This study was performed to assess the immediate and long-term patency of stent-associated side branches (SB) according to the types of stent. A total of 314 patients with 332 lesions (CrossFlex stent 86, NIR 100, GFX 146) had 365 SB (>1 mm) covered by coronary stents. Side branch occlusion (SBO) occurred in 7.7% of CrossFlex stent, in 10.5% of NIR stent and in 8.8% of GFX stent (P = NS). SBO primarily occurred in SB with ostial disease, and the presence of SB ostial disease was the only independent predictors of SBO after stenting (OR 22.1, 95% CI 9.47-51.49, P < 0.001). At 6 months follow-up, 11 of 31 SBO regained the patency, but the remaining SB had persistent SBO. Delayed SBO occurred in 8 SB, being associated with the presence of SB ostial disease and in-stent restenosis. In conclusions, SBO was not associated with the types of stent design, but with the SB lesion morphology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.