Abstract

To experimentally study the effects of altering implant length, outer diameter, cortical bone thickness, and cortical bone density on the primary stability of orthodontic miniscrew implants (MSIs). Maximum insertion torque (IT) and pullout strength (POS) of 216 MSIs were measured in synthetic bone with different cortical densities (0.64 g/cc or 0.55 g/cc) and cortical thicknesses (1 mm or 2 mm). Three MSIs were evaluated: 6-mm long/1.75-mm outer diameter, 3-mm long/1.75-mm outer diameter, and 3-mm long/2.0-mm outer diameter. To test POS, a vertical force was applied at the rate of 5 mm/min until failure occurred. The 6-mm MSIs displayed significantly (P < .001) higher IT and POS than the 3-mm MSIs did. The 3-mm MSIs with 2.0-mm outer diameters showed significantly higher (P < .001) IT and POS than the 3-mm MSIs with 1.75-mm outer diameters. The IT and POS were significantly (P < .001) greater for the MSIs placed in thicker and denser cortical bone. Both outer diameter and length affect the stability of MSIs. Increases in cortical bone thickness and cortical bone density increase the primary stability of the MSIs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.