Abstract

Soil samples were prepared for multi-element analysis using HNO3 leaching or pseudo-total digestion with HNO3, HCl and HF in a microwave oven, both methods requiring 70 min heating time. Two calibration approaches for the soil characterization were also compared: external calibration, combined with internal standardization, and isotope dilution (ID) after appropriate spiking of the soils with a stable isotope mixture prior to sample preparation. Analyses were performed using inductively coupled plasma sector field mass spectrometry (ICP-SFMS). Accurate total elemental concentrations were only obtained for Cd and P using both sample preparation methods in two certified reference materials, NIST SRM 2709 and CCRMP SO-2, as well as comparable values for a Finnish inter-laboratory soil. The pseudo-total digestion method also provided accurate results for As, Be, Co, Fe, Mn, Ni, Pb, Sb, Ti, V and Zn. For Cu in SO-2 and Cr in both certified reference materials, incomplete recoveries were always obtained. In the case of Cr, this is due to difficulties associated with the complete solubilization of refractory minerals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.