Abstract

Background and Aim: Auditory steady-state response (ASSR) has widespread usage in the clinical hearing assessment of infants and young children. The present study evaluated the effect of restricting maximum possible intensity to 100 dB HL on ASSR thresholds. Methods: This retrospective study was conduc­ted in multiple audiology centers. ASSR thre­sholds with restricted settings were evaluated in 58 infants (116 ears). They had absent otoaco­ustic emissions and click-evoked auditory bra­instem response waveforms and bilateral severe to profound hearing loss in behavioral evalu­ations. Results: ASSR thresholds were absent in 28%, 25%, 60%, and 70% of ears in 500, 1000, 2000, and 4000 Hz, respectively. The mean value of remained thresholds was approximately 90−95 dB HL. Conclusion: Restricted ASSR settings are pre­valent; however, they fail to provide extensive additional information about hearing sensitivity. Thus, manufacturers are suggested to improve ASSR setting and transducers and resolve the intensity restrictions. ASSR settings should be able to test auditory thresholds to level of 120 dB HL.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.