Abstract

Finite-element analysis. Intraoperative contouring of rods is a common procedure for spine surgeons to match the native curvature of the spine, but it may lead to premature weakening of the rod. This study investigated the effect of different bending methods on rod fatigue performance. Rod failure in the cervical spine is of clinical concern, particularly when spanning the cervicothoracic region and when considering corrective osteotomies for deformity correction and global spinal alignment. Finite-element models were developed to simulate rod bending (3.5mm D, 40mm L) to achieve a 23° angle with 3 different bending methods: French single, multiple bending, and in situ bending. Simulations were conducted in 4 steps: rod bending, rod spring back, residual stress relaxation, and F1717 mechanical test simulation. French single bending resulted in the highest residual stress concentrations for both titanium (TiAlV) and cobalt chrome (CoCr) at 783MPa and 507MPa, respectively. During F1717 test simulation, the French single bent rod had its highest tensile stress in the middle, with 917MPa and 623MPa, respectively, for TiAlV and CoCr, compared to in situ (580MPa and 586MPa for TiAlV and CoCr) and the French multiple bent rod (765MPa and 619MPa for TiAlV and CoCr). The computational model found that CoCr rods made the construct least prone to deformation. French single bend with TiAlV rods put the construct at highest risk of failure. CoCr rods led to minimal physical changes in microstructure while showing evidence of flattening.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call