Abstract

Osteolysis caused by wear particles from polyethylene in the artificial hip joints is a serious issue. We have used photo-induced radical graft polymerization to graft 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer onto the surface of cross-linked polyethylene (CLPE-g-MPC) in order to reduce friction and wear at the bearing surface of the joint. The physical and mechanical properties of CLPE and CLPE-g-MPC were not significantly different, expect that the friction coefficient of untreated CLPE cups was 0.0075, compared with 0.0009 for CLPE-g-MPC cup, an 88% reduction. After 3.0 x 10(6) cycles in the hip joint simulator test, we could not observe any wear of CLPE-g-MPC cups. We concluded that the advantage of photo-induced radical graft polymerization technique was that the grafted MPC polymer gave a high lubricity only on the surface and has no effect on the bulk properties of the CLPE substrate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.