Abstract

This study aimed to identify the optimal conditions for delivering sufficient doses to deep-seated lesions within short irradiation times for two boron carriers of different T/N ratios. The therapeutic depth and irradiation time of a neutron beam for beam shaping assemblies (BSAs) with a Li or Be target and a MgF2 or CaF2 moderator were examined with the fast-neutron dose per epithermal neutron (FNR) as a parameter. When T/N = 3.61, the therapeutic depth was almost saturated at an FNR of about 10 × 10–13 Gy·cm2; when the FNR value was about 10 × 10–13 Gy·cm2, the therapeutic depth of the neutron beam for the BSA with a Be target and a MgF2 moderator was almost identical to that for the neutron beam for the BSA with a Be target and a CaF2 moderator, and slightly greater than those for the neutron beams for the BSAs with a Li target and a MgF2 or CaF2 moderator; moreover, the irradiation time of the neutron beam for the BSA with a Be target and a MgF2 moderator was shorter than that for the neutron beam for the BSA with a Be target and a CaF2 moderator. When T/N = 100, the therapeutic depths of the neutron beams for the BSAs varied greatly depending on the FNR, and were greater than the corresponding values for T/N = 3.61. We therefore concluded that the BSA with a Be target and a MgF2 moderator that produced a neutron beam with an FNR of about 10 × 10–13 Gy·cm2 is optimal for delivering sufficient doses to deep-seated lesions in short irradiation times when T/N = 3.61, and stricter control over FNR is required when T/N = 100.