Abstract
To confirm that the dimming luminance change in the CRT monitor will have a detrimental effect on diagnostic performance, we performed the four-alternative forced choice experiments on the CRT monitor with the 11 luminance conditions simulating CRT degraded by long-term usage. Six radiologists and one pulmonary physician performed a detection task for 11,000 test areas on 110 test images. The 11 monitor conditions significantly affected the correct detection rates of simulated nodules in the four-alternative forced choice experiments. The gradual deterioration in CRT monitor luminance will have certainly a detrimental effect on detection performance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Computerized Medical Imaging and Graphics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.