Abstract

Therapeutic serum concentrations of metronidazole will lead to under-estimation of ASAT activity when a SMA 12/60 Auto Analyser is used. This interference is due to metronidazole entering the colorimeter through the dialyser membrane in the SMA 12/60 Auto Analyser. This drug is one of the few with a relatively high therapeutic serum concentration, absorbing light in the 340 nm range. Metronidazole will similarily affect other 340 nm methods (LDH, ALAT) available on the SMA 12/60. This interference with metronidazole on the SMA 12/60 can easily be solved by the introduction of a separate serum blank channel for the 340 nm methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.