Abstract

Immunoisolation using semipermeable membranes has been incorporated into bioartificial liver (BAL) devices to separate cellular components of the recipient's immune system from the cells within the BAL device. This study was designed to explore the influence of membrane molecular weight cutoff on performance of the multilayer radial-flow BAL using porcine hepatocytes cocultured with mesenchymal stem cells. In this study, healthy beagles underwent 6-h treatment with a BAL containing membrane with 200 kDa retention rating or 1200 kDa retention rating. Functional markers of BAL performance were monitored before and after treatment, as well as cytotoxic immune response to BAL therapy. The results showed that hepatocyte performance levels such as albumin secretion, urea synthesis, and viability were all significantly higher in 200 kDa retention rating group compared with the 1200 kDa retention rating group after treatment (P < 0.05). Significant levels of canine proteins were detected in BAL medium from the 1200 kDa retention rating group. Fluorescence microscopy further verified that heavy deposition of canine IgG, IgM, and complement (C3) on coculture cells was obtained after BAL treatment in the 1200 kDa retention rating group. However, only trace deposits of canine immunoproteins were observed on coculture cells obtained from BAL in the 200 kDa retention rating group. Small membrane molecular weight cutoff of the BAL could reduce the transfer of xenoreactive antibodies into the BAL medium and improve the performance of the BAL.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call