Abstract
Matrix effects due to Ca, K, Na, and P on analyte transport rate, line intensities, and plasma excitation conditions were studied using a fused silica capillary thermospray nebulizer. Several emission lines of Ba, Cd, Co, Cr, Cu, Mg, Mn, Ni, Pb, and Zn, differing in their excitation and ionization energies, were investigated. In general, the addition of matrix elements to the analyte solution resulted in a reduction of the analyte emission signal. The greatest reductions were observed for matrices containing calcium and for the analyte emission lines with higher sum of excitation and ionization energies. A decrease in the analyte transport rate was the highest for single potassium matrix and for binary matrices containing calcium. The excitation temperature was not significantly changed in the presence of Na, K, and P, but it was depressed for single and composite matrices containing Ca. It was shown that variations in the analyte emission intensities in the presence of concomitants could be attributed to changes in the plasma as well as to changes in the analyte transport efficiency.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.