Abstract

IntroductionManual hyperinflation (MH), a maneuver applied in mechanically ventilated patients to facilitate secretion removal, has large variation in its performance. Effectiveness of MH is usually evaluated by its capacity to generate an expiratory flow bias. The aim of this study was to compare the effects of MH—and its resulting flow bias—applied according to clinical practice versus according to expert recommendation on mucus movement in a lung model simulating a mechanically ventilated patient.MethodsTwelve physiotherapists were asked to apply MH, using a self-inflating manual resuscitator, to a test lung as if to remove secretions under two conditions: according to their usual clinical practice (pre-instruction phase) and after verbal instruction to perform MH according to expert recommendation was given (post-instruction phase). Mucus simulant movement was measured with a photodensitometric technique. Peak inspiratory flow (PIF), peak inspiratory pressure (PIP), inspiratory time (TINSP), tidal volume (VT) and peak expiratory flow (PEF) were measured continuously.ResultsIt was found that MH performed post-instruction delivered a smaller VT (643.1 ± 57.8 ml) at a lower PIP (15.0 ± 1.5 cmH2O), lower PIF (38.0 ± 9.6 L/min), longer TINSP (1.84 ±0.54 s) and lower PEF (65.4 ± 6.7L/min) compared to MH pre-instruction. In the pre-instruction phase, MH resulted in a mean PIF/PEF ratio of 1.73 ± 0.38 and mean PEF-PIF difference of -54.6 ± 28.3 L/min, both out of the range for secretion removal. In the post-instruction phase both indexes were in the adequate range. Consequently, the mucus simulant was moved outward when MH was applied according to expert recommendation and towards the test lung when it was applied according to clinical practice.ConclusionsPerformance of MH during clinical practice with PIF higher than PEF was ineffective to clear secretion in a lung model simulating a mechanically ventilated patient. In order to remove secretion, MH should result in an adequate expiratory flow bias.

Highlights

  • Manual hyperinflation (MH), a maneuver applied in mechanically ventilated patients to facilitate secretion removal, has large variation in its performance

  • It was found that MH performed post-instruction delivered a smaller VT (643.1 ± 57.8 ml) at a lower peak inspiratory pressure (PIP) (15.0 ± 1.5 cmH2O), lower Peak inspiratory flow (PIF) (38.0 ± 9.6 L/min), longer TINSP (1.84 ±0.54 s) and lower peak expiratory flow (PEF) (65.4 ± 6.7L/min) compared to MH pre-instruction

  • Effects of manual hyperinflation on displacement of mucus simulant outward when MH was applied according to expert recommendation and towards the test lung when it was applied according to clinical practice

Read more

Summary

Objectives

The aim of this study was to compare the effects of MH—and its resulting flow bias—applied according to clinical practice versus according to expert recommendation on mucus movement in a lung model simulating a mechanically ventilated patient. The aim of the present study was to compare the effects of MH—and its resulting flow bias—applied according to clinical practice versus according to expert recommendation on mucus movement in a lung model simulating a mechanically ventilated patient

Methods
Results
Discussion