Abstract

The purpose of the current study was to characterize the effects of transvenous ICD shocks on myocardial impedance. Rather than recording impedance during shocks, it was measured during continuous pacing in order to minimize confounding effects such as electrode polarization. Pacing impedance (reflecting the combined impedances of the electrode-tissue interface, myocardium, and blood pool) was measured every 5 seconds before and after 58 single shocks in 22 patients undergoing ICD implantation with a Transvene (n = 14) or Endotak (n = 8) lead. There was a progressive and long-lasting decrease in impedance after shocks. The magnitude of this change was similar for 0.6-J test shocks and shocks > or = 5 J (28 +/- 32 omega vs 23 +/- 16 omega; P = 0.8). However, the drop in impedance was more abrupt after high energy shocks. Because impedance continued to decline throughout the 5-minute interval between shocks, successive shocks had a cumulative effect, with a decrease of 46 +/- 42 omega after four discharges. In conclusion, a progressive decline in pacing impedance is a characteristic response to transvenous ICD discharges.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.