Abstract

Fungal laccase was immobilized on carbon-fiber electrodes using classical methods: physical adsorption, glutaraldehyde, carbodiimide and carbodiimide/glutaraldehyde. The highest biosensor response was obtained using carbodiimide/glutaraldehyde for coupling laccase to carboxyl groups on the carbon fibers. In this method, different percentages of glutaraldehyde had important effects on the sensitivity of the biosensor, the best percentage of glutaraldehyde being 10% (m/v). The behavior of the obtained biosensor was investigated in terms of sensitivity, operational range, pH and applied potential. The developed biosensor showed an optimum response at pH 5.0 and at an applied potential of −100 mV. The immobilized laccase retained a good activity for over 2 months.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.