Abstract
To assess the effect of forced convection of heated air exchange in preterm infants in conventional incubators, we measured insensible water loss and total heat loss in preterm infants in a conventional forced convection incubator (air velocity 15 to 25 cm/second) and in a specially constructed still-air incubator (air velocity 0 to 2 cm/second) at equal operative temperature and humidity. Under the forced conditions, insensible water loss in the preterm infants increased by a mean 52% from 1.04 +/- 0.24 (mean +/- SD) to 1.58 +/- 0.51 ml/kg/hour (P less than 0.001). The ensuing increase in evaporative heat loss was partly reflected in the small but significant increase in total heat loss from 1.65 +/- 0.47 to 1.80 +/- 0.44 kcal/kg/hour (P less than 0.02). In the forced convection incubator, the increased evaporative heat loss in preterm infants was apparently partly compensated by a decreased nonevaporative heat loss. If reduction of insensible water loss is required, preterm infants should not be subjected to forced convection in incubators.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.