Abstract

Bioimpedance analysis was used to measure the residual limb fluid volume of seven transtibial amputee subjects using elevated vacuum sockets and nonelevated vacuum sockets. Fluid volume changes were assessed during sessions with the subjects sitting, standing, and walking. In general, fluid volume losses during 3 or 5 min walks and losses over the course of the 30 min test session were less for elevated vacuum than for suction. Numerous variables, including the time of day that data were collected, soft tissue consistency, socket-to-limb size and shape differences, and subject health, may have affected the results and had an equivalent or greater effect on limb fluid volume compared with elevated vacuum. Researchers should well consider these variables in the study design of future investigations on the effects of elevated vacuum on residual limb volume.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.