Abstract

A series of studies has been undertaken to determine the susceptibility of the liquid sampling–atmospheric pressure glow discharge (LS–APGD) atomic emission source to easily ionizable element (EIE) effects. The initial portions of the study involved monitoring the voltage drop across the plasma as a function of the pH to ascertain whether or not the conductivity of the liquid eluent alters the plasma energetics and subsequently the analyte signal strength. It was found that altering the pH (0.0 to 2.0) in the sample matrix did not significantly change the discharge voltage. The emission signal intensities for Cu(I) 327.4 nm, Mo(I) 344.7 nm, Sc(I) 326.9 nm and Hg(I) 253.6 nm were measured as a function of the easily ionizable element (sodium and calcium) concentration in the injection matrix. A range of 0.0 to 0.1% (w/v) EIE in the sample matrix did not cause a significant change in the Cu, Sc, and Mo signal-to-background ratios, with only a slight change noted for Hg. In addition to this test of analyte response, the plasma energetics as a function of EIE concentration are assessed using the ratio of Mg(II) to Mg(I) (280.2 nm and 285.2 nm, respectively) intensities. The Mg(II)/Mg(I) ratio showed that the plasma energetics did not change significantly over the same range of EIE addition. These results are best explained by the electrolytic nature of the eluent acting as an ionic (and perhaps spectrochemical) buffer.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.