Abstract

Effects of sterilization by electron beam (E-BEAM) on paclitaxel (1) mixed with poly(DL-lactide-co-glycolide) (PLG) in reservoirs of COSTAR Stents are examined by using liquid chromatography-mass spectrometry (LC-MS-MS) techniques with information-dependent acquisition (IDA). Numerous degradation products of 1 are formed in a β-radiation dose-dependent manner to give plethora of low-level degradants. This behavior, together with multiple interferences from PLG-related compounds, creates considerable challenges for analysis of the drug/PLG mixtures. IDA methods with different survey scans are proven to be very efficient in elucidating degradation pathways and in identifying numerous products. Combined LC-MS-MS results from analysis of sterilized drug substance and stents indicate that water addition and oxidative processes together with the isomerization are largely responsible for degradation of 1 under E-BEAM sterilization conditions used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.