Abstract

To evaluate the effect of processing solution depletion on the image quality of F-speed dental X-ray film (Insight), compared with Ektaspeed Plus. The films were exposed with a phantom and developed in manual and automatic conditions, in fresh and progressively depleted solutions. The comparison was based on densitometric analysis and subjective appraisal. The processing solution depletion presented a different behaviour depending on whether manual or automatic technique was used. The films were distinctly affected by depleted processing solutions. The developer depletion was faster in automatic than manual conditions. Insight film was more resistant than Ektaspeed Plus to the effects of processing solution depletion. In the present study there was agreement between the objective and subjective appraisals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.