Abstract

Dissolution testing is a critical method for the determination of pharmaceutical product quality and bioequivalence. For some products, dissolved gases in the dissolution medium affect dissolution results thus requiring degassing of the medium prior to use. In this study, we use a total dissolved gas and oxygen meter to measure both oxygen and total gases in dissolution media before and after application of a variety of deaeration methods. Dissolution testing results using a 10 mg Prednisone tablet (NCDA #2) are compared with the percent saturation of oxygen and total gases found in the medium. Reaeration of the medium during different stirring rates was also measured. This study confirms that measurement of total gases and not just oxygen in the medium is necessary to assess adequacy for dissolution testing. For those deaeration techniques that are performed at room temperature, the percent saturation of the total dissolved gases must be well below 100% to prevent outgassing once medium is brought to dissolution test method temperature, typically 37 degrees C.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.