Abstract

AimsMyocardial ischemia can affect traditional right ventricular (RV) pacing parameters, but it is unclear whether coronary artery disease (CAD) impact the pacing parameters and electrophysiological characteristics of left bundle branch area pacing (LBBaP) as a physiological pacing representative. MethodsPatients who underwent coronary angiography (CAG) after/before the LBBaP procedure and underwent percutaneous coronary intervention after LBBaP procedure were divided into CAD group and Non-CAD group according to visual CAG. Pacing parameters and electrophysiological characteristics were recorded at LBBaP implantation. Multivariate logistic regression analysis was implemented to evaluate the association between CAD and higher capture threshold. Sensitivity analyses were conducted to verify result stability. ResultsA total of 176 patients met inclusion criteria (115 Non-CAD patients and 61 CAD patients) with a mean age of 71.1 ± 9.0 years. Compared with the Non-CAD patients, CAD patients had the higher capture threshold (0.67 ± 0.22 V vs. 0.82 ± 0.28 V, P < 0.001) and lower R-wave amplitude (12.5 ± 4.8 mV vs. 10.1 ± 2.7 mV, P = 0.001). Moreover, CAD was independently associated with higher capture threshold (adjusted Odds ratio (OR) 3.418, 95% confidence interval (CI): 1.621–7.206, P = 0.001), which was further validated through sensitivity analyses. ConclusionPatients without CAD might have safer pacing parameters in the LBBaP procedure. Besides, CAD might be the risk factor of capture threshold increase during permanent LBBaP implantation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call