Abstract

The effects of the column length, the particle size, the gradient length and the flow rate of a nanoLC system on peptide peak capacity were investigated and compared. Columns packed with 1.7 μm and 3 μm C 18 materials into pieces of 75 μm capillary tubing of various lengths were tested with different gradient lengths and flow rates. While increasing the length of a column packed with the 1.7 μm material helped improve peptide peak capacity at the whole range of the tested gradient lengths (24–432 min), little improvement in peak capacity was observed with the increase of the length of a column packed with the 3 μm material unless a gradient longer than 50 min was carried out. Up to 30% of peak capacity increase was observed when a column's length is doubled, with little reduction in the throughput. In most cases, more than 50% of the increase in peak capacity was obtained with the reduction in the particle size from 3 μm to 1.7 μm. With the same backpressure generated, a shorter 1.7-μm-particle column outperformed a longer column packed with the 3 μm material. In a flow rate range of 100–700 nl/min, increasing the flow rate improved peak capacity for columns packed with 1.7 μm and 3 μm materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.