Abstract

The purpose of this article is to review the MR principles of blood signal and normal flow-related phenomena seen in the aorta and inferior vena cava (IVC) on T2-weighted images of the abdomen and to discuss the physiologic and physical basis of signal alterations with breath-hold imaging. We define time-of-flight (TOF) loss and list the factors that affect it; explain the physiologic effects of breath-hold imaging on both aortic and IVC waveforms and velocity; state which abdominal T2 imaging techniques are most susceptible to variable TOF effects and explain why based on physiologic effects and MR principles; and describe three trouble-shooting techniques to confirm that unexpected signal does not reflect pathology. T2 images customarily result in TOF loss and dark blood signal. Abdominal MRI relies extensively on breath-hold imaging techniques that may alter markedly the flow velocity in normal vessels. This marked flow change results in signal heterogeneity and variable TOF loss especially with faster sequences filling k-space in shorter time frames. Breath-hold imaging decreases k-space fill time, and of the breath-hold T2 techniques, single-shot fast spin-echo decreases k-space fill time the most. Atypical blood signal during breath-hold imaging may mimic pathology, but abnormal findings can be verified by review of other sequences and planes of imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.