Abstract
The BAHA® Softband has been developed to provide a transcutaneous anchor for a BAHA® until a child is a surgical candidate for the percutaneous BAHA® implant. We tested the objective output force level of the BAHA® Classic 300 and Compact connected to a Softband on an artificial mastoid to determine: (1) the effects of direct contact force on output force levels (dB); and (2) the required volume control setting to ensure audibility of speech (assuming an average adult reference equivalent threshold force level).Direct contact force was varied from 2 to 5 N in 1 N steps. Output force level increased with increasing contact force. However, the average increase was 3 dB or lower, suggesting that the contact force is of minor importance. Volume control setting appears to be of much greater importance. It is suggested that the volume setting of either device be set to at least 2.5 to ensure audibility of conversational speech. Data from normal-hearing adults with simulated conductive hearing losses are presented to validate this conclusion.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.