Volumetric capnography in healthy ventilated neonates showed deformed waveforms, which are supposedly due to technological limitations of flow and carbon dioxide sensors. This bench study analyzed the role of apparatus dead space on the shape of capnograms in simulated neonates with healthy lungs. We simulated mechanical breaths in neonates of 2, 2.5, and 3 kg of body weight using a neonatal volumetric capnography simulator. The simulator was fed by a fixed amount of carbon dioxide of 6 mL/kg/min. Such simulator was ventilated in a volume control mode using fixed ventilatory settings with a tidal volume of 8 mL/kg and respiratory rates of 40, 35, and 30 breaths per minute for the 2, 2.5 and 3 kg neonates, respectively. We tested the above baseline ventilation with and without an additional apparatus dead space of 4 mL. Simulations showed that adding the apparatus dead space to baseline ventilation increased the amount of re-inhaled carbon dioxide in all neonates: 0.16 ± 0.01 to 0.32 ± 0.03 mL (2 kg), 0.14 ± 0.02 to 0.39 ± 0.05 mL (2.5 kg), and 0.13 ± 0.01 to 0.36 ± 0.05 mL (3 kg); (p < .001). Apparatus dead space was computed as part of the airway dead space, and therefore, the ratio of airway dead space to tidal volume increased from 0.51 ± 0.04 to 0.68 ± 0.06, from 0.43 ± 0.04 to 0.62 ± 0.01 and from 0.38 ± 0.01 to 0.60 ± 0.02 in the 2, 2.5 and 3 kg simulated neonates, respectively (p < .001). Compared to baseline ventilation, adding apparatus dead space decreased the ratio of the volume of phase III to VT size from 31% to 11% (2 kg), from 40% to 16% (2.5 kg) and from 50% to 18% (3 kg); (p < .001). The addition of a small apparatus dead space artificially deformed the volumetric capnograms in simulated neonates with healthy lungs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call